INSTRUCTIONS LOW FLOW KIT R479
Kit R479 is a collection of parts and tools required to adapt a standard FMI stainless steel pump head to
the requirements of low flow (less than 50 ml per minute) pumping applications where reduction of system
dead volume and rapid fluid transit time are important and the use of small bore tubing is indicated. The
kit serves to adapt from the ¼” NPT internal threads of the standard stainless steel cylinder case to the
¼-28 internal machine threads required by most commercial small bore tube fittings.
Kit R479 consists of four ferrules R478 (one for each port of the pump head and two spares), two
adapters #110339 (one for each port), one ferrule removal tool (screw) and one ferrule assembly press
tool (dowel).
A. To Install Low Flow Components:
1. Remove tube adapter R412 from standard head assembly; flush away Teflon tape shreds and other
residue that may remain in the port threads of the pump head.
2. Press one ferrule R478 into the small center hole of each side port, using press tool (dowel) to seat
each firmly and squarely on the slotted outer surface of port plug R419.
3. With piston assembly R423 bottomed in the cylinder and piston pin #110301 at 6 o’clock position,
screw adapter #110339 into each port, hand tight against the ferrule plus 1/8 turn wrench tight. Do not
over tighten these fittings as excessive pressure on cylinder liner R407 may cause it to bind on the
piston R423.
4. Connect tubing using ¼-28 commercial fittings (supply house item) and operate pump to pump fluid
against 5 to 10 psig head resistance for several minutes. Check for leaks. If fluid shows around the
threads of the adapter #110339 that is operating under pressure, tighten it 1/16 to 1/8 turn more. If
the leak persists, remove kit parts and start over, using greater care in axially aligning ferrule R478 in
the center hole of port plug R419. Reverse the fluid flow direction and test the other port under
pressure.
B. To Remove Ferrule #R478:
Screw the removal tool (sheet metal screw) into the center hole of the ferrule by hand until the ferrule
can be pulled out with the screw.
IN-R479-10
R479 KIT PARTS INSTALLED
CYL. LINER R407
R478 FERRULE
110339 ADAPTER
R419 PORT PLUG
R405
CROSS-SECTION VIEW OF STANDARD STAINLESS- STEEL CYLINDER
CASE ASSEMBLY AS ADAPTED FOR LOW FLOW, LOW DEAD VOLUME
APPLICATIONS
Page view 0
1

Summary of Contents

Page 1 - APPLICATIONS

INSTRUCTIONS LOW FLOW KIT R479 Kit R479 is a collection of parts and tools required to adapt a standard FMI stainless steel pump head to the requi

Comments to this Manuals

No comments